Sound and noise are simply pressure waves
traveling through the air at the speed of sound (1122 ft/sec or 765 mph)
under standard atmosphere conditions. (Sound waves can also travel through
other media at higher velocities, but that is outside the subject of
transducers.) It logically follows that to measure sound it is only necessary
to measure the pressure of the wave(s) caused, using a suitable transducer .
The actual physical pressures involved are quite small but ex tend over
a considerable range. Sound pressure levels are expressed in decibels
(dB), the significance of which will be explained later. The actual pressure
produced by any sound pressure level is then given by
pressure (lb/in.^2) = 29 x 10^dB/20 x 10^-10
For example, 80 dB is a fairly loud sound. Calculating the equivalent
pressure, we get
pressure = 29 x 10^80/20 x 10^-10
= 29 x 10^4 x 10^-10
= 0.000029 lb/in
We need a very sensitive transducer to respond to such low pressure
levels and connect it into a measurable/undetectable quantity (for example,
a meter reading). Fortunately, we have one in the microphone. Microphones
are, in fact, the standard form of transducer used for measurement of
sound.
Now to return to the subject of decibels. A human being (and presumably
also animals, birds, and other creatures) detect sound as an energy intensity
level. The lowest sound intensity level that can be detected by the average
individual is 10^-12 W/m^2. This level is described as the threshold
of hearing
With increasing intensity levels, sound appears louder and louder, until
eventually the sensation received changes from hearing to feeling. This
occurs at a sound intensity level of about 1 W/m and represents the threshold
of feeling
The intensity or sound energy range covered between just hearing and
the threshold of feeling is thus from 10^-12 to 1, or a range of 1 million
million. Such a simple linear scale is far too large to work with, so
we use one related to the ratio of intensity levels, called bels, and
expressed by the formula
bels = log10 (Ih/Io)
where Ih is the sound intensity level of the sound heard
Io is the sound intensity level at the threshold of hearing
Even this proved too broad a scale, so a smaller unit of one- tenth
of a bel, called a decibel, is used: thus,
decibels (dB) = 10 log10 (Ih/Io)
This gives a range of 120 dB between the threshold of audibility and
the threshold of feeling. It also establishes a useful figure to remember:
a doubling in sound intensity corresponds (almost exactly) to a rise
of 3 dB.
This can be proved as follows. If I represents the original sound intensity,
and 12 the doubled sound intensity, the dB relationship is
dB = 10 log10 (I2/I1)
= 10 log10 (2I1/I1)
= 10 log10 (2/I1)
= 10 log10 (2/I1)
= 3.01
This is not the complete story, especially concerning sound transducers.
They detect sound pressure level. Now sound intensity is proportional
to the square of the corresponding sound pressure level, so the dB relationship
for sound pressure levels becomes
dB = 20 log10 (P2/P1)
In other words, doubling the sound pressure level is equivalent to a
rise of 6 dB.
Thus because sound is measured in terms of sound pressure level, a change
of 6 dB represents a doubling or halving of sound intensity as distinguished
by the human ear.
BASIC SOUND MEASUREMENT TECHNIQUE
A microphone converts sound pressure impinging on it into an electrical
signal that can be amplified as necessary to provide a readout. However,
the transducer signal output will normally be a ragged ac-type signal
alternating rapidly from positive to negative levels, which is impossible
to display except on an oscilloscope. To obtain a meaningful readout
for instrumentation, therefore, you must interface the transducer with
circuitry that modifies the out put signal so that both positive and
negative parts are rendered as a positive signal, although varying in
amplitude at audio frequencies. For a steady “average” reading the signal
is put through a time constant circuit; the square root of the average
signal is then extracted to give a root-mean-square (rms) average reading.
In the practical sound-level meter the time constant circuit normally
provides two averaging time constants: one slow (about 0.9 sec), and
one fast (about 0.125 sec). Slow speed is selected for measuring fluctuating
sounds; fast speed is selected for measuring continuous, reasonably steady
noise. If in doubt as to the character of the noise, you can measure
it on both settings and com pare them. If the fluctuations in meter reading
on fast-setting range are less than 6 dB, then the average reading should
agree with the reading attained on slow setting.
In addition, the sound-level meter incorporates a calibrated alternator
network, enabling reading range to be switched to various levels (usually
in 10-dB steps). This basically is equivalent to ex tending the length
of scale of the meter reading by the number of alternative steps provided.
WEIGHTING NETWORKS

Fig. 22-1. Relative sound pressure required to produce a given perceived
level of sound as a function of frequency.
Averaged sound pressure levels read as a single number indicated on
a meter still have one basic limitation. They are a purely quantitative
measurement of sound, which differs from subjective response to sound
level at different frequencies (see Fig. 22-1). Here the curves correspond
to different “loudness levels,” as heard, showing how apparent loudness
changes with frequency. It will be seen that at lower frequencies lower
sound pressure levels give the same apparent loudness. Thus, low-frequency
sounds appear to be “louder” than higher-frequency sounds at the same
sound pressure level. (High-frequency sounds can be more disturbing,
however.)
To overcome this, we add a weighting network to the meter circuit, which
has the effect of adjusting the final averaged signal out put to compensate
for this effect (Fig. 22-2). The usual form of employment is A weighting,
when the corresponding scale readout is calibrated in dB(A). This is
now universally employed for single- number noise measurement and is
generally in close agreement with subjective ratings. Meters may also
incorporate other weights, such as B, C, and D, but those are used for
more specialized purposes.

Fig. 22-2. An approximate graph of relative gain versus frequency
for an equalizing circuit, designed so that the human ear perceives
sound intensity according to the actual sound pressure.
The single-number reading dB(A) sound-level meter is a relatively simple
instrument, easy to use and understand, and generally suitable for comparative
purposes. It still has many limitations, however. It is measuring only
a (weighted) average sound pressure level that gives no indication of
the active frequency content of the sound. More complex instruments employ
additional filters and circuitry to measure sound pressure levels at
different frequencies, which enables the actual distribution of sound
content to be analyzed; from this a frequency spectrum can be plotted
or presented on a display. This is concerned entirely with electronic
circuitry and is outside the subject of transducers. The transducer (microphone)
provides the original signal information only. To what degree it is amplified,
averaged, weighted, alternated, and analyzed depends on the following
electronic circuitry and readout system adopted.
MICROPHONES
Microphones are the only transducer involved in noise- measuring instruments.
The two main types are piezoelectric and condenser. Piezoelectric microphones
employ a crystal of piezoelectric material as a transducer. They may
also be known as c or ceramic microphones. Condenser microphones, also
known as electrostatic or capacitor microphones, work on the principle
of variations in electrical capacitance to provide an output signal.
They are generally more prone to self-noise and require more extensive
instrumentation than ceramic types, but they can offer advantages for
specific applications—for example, when good high-frequency characteristics
are required.
Other types of microphones that may be used include moving- coil or
dynamic types and special directional microphones, which may incorporate
any one of the three types of elements. Dynamic microphones are, however,
relatively limited in application and have the basic disadvantage that
comparatively large sizes are necessary to avoid distortion at low frequencies.
Hydrophones are a further type of microphone, designed specifically
for underwater sound measurement. They normally employ piezoelectric
elements.
Table 22-1 gives a general comparison of different types.
(i) Piezoelectric microphones (a) quartz
crystal
(b) ADP
(c) barium titanate
(d) lead zirconate
(e) ceramic
(ii) Condenser microphones
crystal
(iii) Moving coil
(iv) Hydrophones
|
response, linear sensitivity, very
good stability, very good remarks—very rugged type of transducer
response, linear sensitivity, very good stability, very good
response, linear sensitivity, fair to good stability, good remarks—can
be affected by water response, linear sensitivity, good stability,
very good remarks--very rugged type of transducer
response, linear sensitivity, good to very good stability, good
to excellent remarks—specially treated ceramic piezoelectric
crystals are now usual choice with good, stable performance at
economic price
response, linear
sensitivity, good stability, very good to excellent remarks--clamped
or stretched diaphragm
construction, electrostatic or capacitor modes. The latest types
are electric foil microphones (capacitor type) permanently polarized
with an electrostatic charge during manufacture. Condenser microphones
are more prone to self-noise.
response, failing at low and high frequencies
sensitivity, good stability, very good
remarks—response tends to be poor at low frequencies.
ceramic crystal type, used for underwater sound measurement. |
The choice of microphone depends on a number of factors. Low self-noise
is obviously desirable, for example; for a microphone used to measure
low sound levels, this is essential. The ceramic type is the preferred
choice in this case, because even under the best conditions the inherent
self-noise of condenser-type microphones can seldom be reduced below
about 20 dB. A further requirement for the measurement of low sound levels
is that the output voltage generated by the microphone must override
the circuit noise of the amplifier in the sound-level meter.
The measurement of high sound levels requires that the system be free
from microphonics or spurious signals generated by vibration. This is
generated in the circuit components rather than the microphone, but mechanical
vibration can also affect the microphone itself. A relatively soft type
of mounting and special types of low-sensitivity microphones for measurement
of very high sound levels may be required.
Microphonics are not normally experienced, except where sound pressure
levels in excess of 100 dB are being measured, unless the microphone
or instrument is mounted in such a way that mechanical vibrations are
transmitted directly to it.
Both the piezoelectric and condenser microphones are generally well
suited for the measurement of low-frequency noise, although moving-coil
types can have distinct limitations imposed by their physical size. Accurate
measurement of high-frequency sounds calls for the use of a small-size
microphone. Regarding frequency response, the condenser type offers a
narrower deviation range and the possibility of extending accurate measurements
up to frequencies of 20,000 Hz. Both types can, of course, be calibrated
for correction. Special types of microphones are required for accurate
measurement of frequencies in excess of 20,000 Hz, or above the audio-frequency
range.
The final choice of type can also be affected by humidity and temperature.
A number of piezoelectric crystals, for example, are sensitive to humidity
and can even be damaged if humidity is too high. In unfavorable ambients
the piezoelectric element must therefore be rated as suitable for working
at the prevailing humidity and/or temperature.
Condenser microphones are also sensitive to humidity, although they
are not necessarily damaged by high humidity. Sensitivity, in this case,
is caused by electrical leakage across the microphone, which tends to
increase with increasing humidity. Exposed insulating surfaces are generally
treated to maintain low leakage with high humidity, but it is not necessarily
positive protection under all conditions. Some authorities recommend
that the microphone be kept at a higher-than-ambient temperature to reduce
leakage.
Other microphones have much higher maximum service temperatures. Condenser
microphones normally have a maximum service temperature of the order
of 100°C. In the case of portable instruments, temperature limitations
are normally restricted to battery performance. Battery life will tend
to be shortened at temperatures in excess of about 500 C., and battery
output may be insufficient at near-zero or subzero temperatures, depending
on battery type. Special batteries are required for satisfactory operation
at subzero temperatures.
All microphones are temperature sensitive. Calibration is carried out
at a normal ambient temperature, and sensitivity will vary at different
temperatures. This is generally negligible in the case of piezoelectric
microphones, where the temperature coefficient is only of the order of
-0.01 dB per degree Celsius. With a typical condenser microphone the
temperature coefficient can be as much as four times greater.
Fig. 22-2. An approximate graph of relative gain versus frequency
for an equalizing circuit, designed so that the human ear perceives
sound intensity according to the actual sound pressure.
MICROPHONE CALIBRATION
Microphones for sound-level meters are calibrated by the manufacturers
with microphone reciprocity calibrators, whereas sound-level calibrators
are used for overall calibration of sound-level meters. Acoustic calibrators
can provide an overall calibration check in field use, but not necessarily
with original accuracy. Many meters have built-in electrical calibration
and internal calibration controls that enable electrical calibration
to be carried out, but this is only a check on the stability of the electrical
system.
The following comments largely cover the question of calibration as
far as the average user of sound-level meters is concerned:
- Initial (manufacturer’s) calibration of the instrument can usually
be regarded as adequate for about a year, after which a calibration check
should be made. Such calibration is normally based on free-field response
to noise of random incidence. The accuracy given by the calibrated meter
then depends on the stability of the microphone and stability of the
electrical system; the validity of the measurement is dependent on the
characteristics of the sound field present when the measurement is taken.
- Electrical calibration is a useful check on the stability of the electrical
system of the calibrator and can be carried out quite simply at regular
intervals, such as monthly (especially if meters have built-in electrical
calibration and internal calibration controls).
- The use of an additional (calibrated) microphone can pro vide a simple
comparative check as to whether the original microphone has changed in
characteristics and thus needs calibration. This method, used with electrical
calibration, provides the simplest overall calibration check. Note, however,
that this depends on the second microphone being used only for checking
and being carefully stored.
- Acoustic calibrators can provide an overall calibration check under
“field” conditions, but the accuracy of such a calibration check depends
both on the suitability of the instrument and its proper method of employment
(particularly regarding sealing of the microphone in the sound chamber).
- An exact calibration check can only be carried out under laboratory
conditions. It is generally recommended that instruments be returned
to the manufacturer(s) at suitable intervals (about once a year) for
such a check.
- Calibration should always be checked if the instrument is subject
to shock, such as being dropped or receiving rough handling.
USING SOUND-LEVEL METERS
Most people using sound-level meters adopt a “point-and-read” technique.
This is generally (but not completely) satisfactory when sound levels
are being measured in a large open area (for example, outdoors), or under
free-field conditions. This broadly implies that there are no surfaces
present to produce sound reflection at the point where the measurement
is being taken. Most meter microphones are omni-directional and thus
pick up sounds coming from all directions.
Wind blowing directly onto the microphone can produce false readings,
although this effect is minimized when the microphone itself is fitted
with a windshield. Note also that wind can affect the strength of noise
measured from a distant source, depending on wind strength and direction
and the distance from the source. The same noise source will give a higher
reading downwind than up wind at the same distance from the source, for
example.
Although most microphones used for sound measurement are omni-directional,
this may only apply as far as low-frequency sounds are carried. At higher
frequencies, where the wavelength of the sound may be comparable to the
size of the microphone itself, considerable directional effect may be
present. In this case the measured response will vary with the direction
in which the microphone is pointed.
As a general rule, for accurate measurement a sound-level meter should
be held at arm’s length sideways with the microphone pointed away from
the noise source, that is, with the sound impinging on the microphone
at grazing incidence (90 deg). However, this will depend on the type
of microphone. Some may need to be pointed directly at the noise source
for free-field measurement. Requirements in this respect will be specified
in the manufacturer’s instructions. Errors as large as ±6 dB may occur
through bad positioning of a hand-held sound-level meter, due mainly
to the presence of the operator. For most accurate results the meter
(or separate microphone) should be clamped in position, and operator
and any other instrumentation moved a short distance away.
MEASURING INDOOR NOISE
Accurate measurement of noise levels in rooms or buildings is complicated
by the fact that sound reflections are inevitably present, the strength
of which will vary at different points. Thus, the actual position at
which the reading is taken is significant. Also if the noise generated
by a particular source is being measured (for example, a machine), the
meter records not only machine noise but background noise as well.
To complicate the question still further, the behavior of noise in the
near field differs from that in the far field; it is also affected by
the shape of the noise field.
Here “field” refers to the hemispherical distance of the point of measurement
from the source—call this R—or the radius of the hemisphere, which is
simply the linear distance between noise source and the point considered.
Normally sound level decreases according to distance squared, that is,
by I/R^2. In the near field, or for small distances (small values of
R), sound intensity will decrease at a lower rate than given by this
law. At some value of R, propagation of sound-level intensity then changes
to follow the inverse-square law, or the condition changes from near
field to far field.
Measurements made in the near field would, therefore, give a false value
if extrapolated on the noise-law basis to estimate such level at a greater
distance. Similarly, measurements taken in the far field would give a
false (low) rate if used to estimate noise level at a short distance
from the noise source in the near field.
The point at which the near field changes to the far field depends both
on the bulk and shape of the noise source and on whether noise is emitted
uniformly from that shape. If we assume that the latter applies, then
the extent of the far field can be estimated as being at least three
times, and no more than five times, the linear dimension of the noise
source “facing” the point considered. Taking measurements at a greater
distance than this should ensure a true far field reading when the noise
law applies for any 5 recalculations or estimates required.
Note, however, that most practical noise sources have both bulk and
non-uniform radiation of sound. In other words, they are directional
noise sources, radiating sound more in some directions than others. This
can only be discovered by taking measurements at various positions around
the source and, if necessary, plotting equal sound-level contours.
In general, sources of low-frequency noise tend to be non-directional
(uniformly radiated), particularly when the source itself is small in
comparison with the wavelength of the sound being generated. Directional
effects tend to become more apparent as the sound frequency increases,
particularly when the source of the sound becomes large in comparison
with the wavelength of the sound.
REVERBERANT FIELD
In a room the walls, ceiling, and floor all form reflecting surfaces.
If we assume a point source of sound, the sound field within the room
has two components:
- The direct sound SD between the source (S) and receiver (R) (Fig.
22-3A). This is the same as free-field conditions and is not modified
by the presence of the enclosure.
- The reflected sound SR, which reaches the receiver after one or more
reflections from the enclosure surfaces (Fig. 22-3B). The value of this
is determined by the power of the source and the reflecting properties
of the enclosure surfaces. With perfect reflection the value of SR would
eventually rise to extreme values, regardless of the value of the source.
In practice, this is impossible, but nevertheless the combined value
of SD and SR can give rise to very high levels at any point R with highly
reflective enclosure surfaces. Marked variations in sound pressure level
at different points R will also be experienced due to interference effects
between the multidirectional waves. This will be further modified by
a non-point source.
If the size of the source is small relative to the dimensions of the
room, there will be a definite distinction between the effects of the
direct sound and the combination of direct and reflected

Fig. 22-3. At A, direct propagation of sound; at B, reflected propagation.
SEMIREVERBERANT FIELD
sounds. Near to the source, direct sound will predominate, and the effect
of reflected sound components will tend to be negligible. At some greater
distances, reflected sound will predominate: the region in which this
occurs is known as the reverberant field. The inner boundaries of this
field will be determined by the size of the room, the absorption characteristics
of the reflecting surfaces in the room, and the acoustic power, size,
and directivity characteristics of the source.
Fluctuations in sound pressure levels in the reverberant field due to
“interference” effects yield patterns that are known as standing waves.
In general, these are most marked when the source has frequency components
corresponding to, or close to, any of the possible resonances of the
room. The spacing of such standing waves tends to be slightly greater
than one-half of the wavelength concerned.
In practice, in most rooms in which sound measurements are semi-reverberant:
the walls, ceiling, and floor are neither completely reflecting nor completely
absorbent. If sound measurements are made in such a room in the same
way as in a free field, correction is necessary to obtain equivalent
free-field sound pressure level readings. This necessitates evaluation
of the room characteristics or room constant. Rooms with a large degree
of absorption have a large room constant and approach free-field conditions.
Rooms with a large degree of reflection and only a small degree of absorption
have a low room constant and approach reverberant-field conditions.
The main constant can be expressed in terms of the ratio of sound decay
in the room (D). This is normally determined as reverberation time or
the time in seconds required for the sound pressure level to fall to
60 dB when the source is shut off. The decay rate then follows as
D= 60 / reverberation time
Once the decay rate has been established or estimated, the average sound
level in that part of the room where the reverberant field applies can
be estimated from the equation
SPL = PWL - 10 log10 V - 10 log10 D + 48
where PWL is the source power level
V is the volume of the room in cubic feet
D is the decay rate, as defined above.
It should be noted that this formula gives approximate answers only.
Correction may also be needed to compensate for changes in atmospheric
pressure, temperature, and relative humidity.
BACKGROUND NOISE
The question of background noise causing inflated readings when measuring
the noise of some individual source is normally not too significant.
Take a reading with the noise source switched off. This is the background
noise. Now take a reading with the noise source switched on. This is
a “combined” reading of the noise source and background noise level.
However, if the difference between the two readings you have taken is
6 dB or more, then the effect of the background can be ignored entirely.
Suppose, for example, the background noise level measured was 55 dB.
With the noise source switched on the new reading is 65 dB. This is a
10-dB difference, so certainly the combination of the background noise
can be ignored. It is only when two nearly equal sound sources are present
at the same time that the lower one has any appreciable contribution
to noise level measured.
Here, however, we must be careful. The combined sound level of two near-equal
noise sources is not the arithmetical sum of the two dB values. The amount
added to the overall noise level is related to the difference between
the two. This is somewhat complicated to work out mathematically, but
Table 22-2 gives satisfactory answers. For differences greater than 6
dB the amount to be added becomes negligible.
Table 22-2. Combining Near-Equal Noise Levels
Difference between sound levels
0dB
1dB
2dB
3dB
4dB
5dB
6dB |
Amount to be added to loudest sound
level
3 dB
2.5 dB
2 dB
1.75dB
1.5 dB
1.25dB
1 dB |
Example 1. Find the combined sound level of two equal sound sources,
each of 55 dB.
difference = 0
so amount to be added is 3 dB
Combined sound level = 55 + 3 = 58 dB
Example 2. Sound level A is 68 dB, and sound level B is 71 dB.
Find the combined sound level.
difference = 3 dB
so amount to be added to loudest (B) is 1.75 dB.
Note. In practice, sound levels are never quoted in decimal fractions
values are rounded up or down to the nearest whole number.
In this case, therefore, the combined sound level could be given as 73
dB.
|