Posted: Tue Nov 27, 2007 12:54 pm Post subject: Maintaining and Recertifying Environmental Test Chambers |
|
|
This guide is also useful when purchasing Used or Refurbished Environmental Test Chambers.
Choose a reputable company -- one with trained and certified technicians -- to purchase a used chamber recertify your existing test chamber.
A checklist of items techs look at and test are:
Refrigeration System Checklist -- inspect the following and repair/replace as necessary:
System refrigerant charges checked to make sure they are within spec..
Compressor oil replaced.
All filter dryers replaced.
Refrigeration valves replaced with rapid-cycle valves unless already existing.
Oil separators replaced unless existing deemed clean.
Pressure switches added for compressor safety unless already existing.
Flush refrigeration system w/ nitrogen to eliminate old oil in lines.
Pressure-test system to find leaks.
Pull vacuum on system to find leaks.
Charge system with new environmentally-friendly refrigerants. E.g. R404a and R508b.
Refrigeration evaporator coils
Condenser unit
Compressors
Electrical Checklist -- inspect the following and repair/replace as necessary:
Contactors, relays
Heaters
Fan motors
Lights, lamps
Wiring, connectors, harnesses
Controllers -- inspect the following and repair/replace as necessary:
Older style controllers without tech. support retrofitted with new controllers
Wiring reconfigured to fit new controller
Controllers tuned for best performance.
Chambers larger than 5 cubic ft. equipped with over temperature-protection controller
General inspections and/or revamp:
All Window
Castors (wheels)
All sheet metal components
Humidity system boilers disassembled + cleaned; Inspect Heating elements + floats
Demineralizer cartridges -- replace.
Door gaskets
Chamber prepped for paint and repainted
Labels applied to chamber as necessary
Chamber re-tested after being painted.
Multiple test profiles run on chamber to ensure reliability. Customer-supplied tests also run upon request Chart results given to customer to verify test results.
Final checkout to ensure that chamber meets original manufacturers specs.
Optional Tests/Services:
Extended warranty
Original compressors replaced with new
Certified calibration
Startup and training
Retrofit from water cooled to air-cooled or visa versa.
Retrofit with accessories such as:
Chart Recorder
Dehumidifier
Feed-Through Ports
Extra Shelves
Water Recycling System
Heat Boost for Faster Temperature Ramp-Up Rates.
LN2 or CO2 Boost for Faster Temperature Ramp-Down
Rates.
Castors if not already existing.
Window installed if not already existing.
Extra event outputs from controller to interface to product
under test.
RS-232 or IEEE-488 Interface to Controller.
GN2 Purge
Be sure to ask about the Warranty of a used/recertified chamber or the work done on your existing chamber.
Period: Commonly 90 days parts + labor within the United States.
Location: Will Warranty work be provided at customer’s location – No need to return equipment unless under certain circumstances.
Is an Extended warranty available? |
|
| Back to top |
|
 |
|
|
You cannot post new topics in this forum You cannot reply to topics in this forum You cannot edit your posts in this forum You cannot delete your posts in this forum You cannot vote in polls in this forum
|
|
Powered by phpBB © 2001, 2005 phpBB Group
|